SP Application
Mammomat Novation DR
Replacing the TFT Monitor
It must be noted that the cabling on the back of the unit must be made in accordance with
the designations.
Selecting the Test Images
The test images are located under "Service\Acquisition System\Quality Tests" with the
designation "Display". Start the service software for this purpose.
•
Select Service and Local Service in the menu bar under Options.
•
Delete the password in the SW screen and click on OK.
Fig. 8:
Display
Mammomat Novation DR -1 Service SW
TD00-000.841.02.04.02
05.05
System-related Adjustments
Page 14 of 22