AX Application

AXIOM Artis

Size and Position
Check the size and position in the norm being used. The picture must be displayed "full
screen".
The emergency mode norm is used "only" as a display; it is displayed in reduced size.
If this is not the case, if needed, the size and position (only position is possible) must be
performed manually.
Frequency / Phase (Focus)
NOTE
Using SMPTE:
Tab. 2
SMPTE
Select the SMPTE test pattern in the
system.
The 5 high-frequency line groups must
be equally well visible!
The adjustment is performed under: =>
Menu => Other => using the sliders:
Frequency and Phase
Using the Text Page (more exactly adjustable)
Select Explorer to specify the background color of Notepad and to select Notepad.
Select Explorer:
Display
This adjustment ensures that the picture is correct for the screen
size.
Incorrect adjustment is manifested as unsharp focus.
The adjustment can be performed using the SMPTE test pattern
(quick adjustment) or by generating a text page.