Repair / Troubleshooting; Test And Auxiliary Equipment - Siemens SCD 1897-M Replacements Of Parts

18" color tft monitor
Table of Contents

Advertisement

General Information
A laptop, etc. is not required; the adjustment is performed using the push-buttons on the
side of the unit and an on-screen menu.
The display can be controlled using RGB as well as S/W.
For configuration for S/W control, see the Instruction Manual.
For technical data, etc., see the Instruction Manual.
Video norm:
- The display detects the video norm that is present and sets itself automatically to this
norm.
- "New" norms that have not yet been set, can (must) be set. The next time this norm is
used, the display sets the norm again automatically.
- There is separate set of parameters in the display for each norm. This means that
each norm used must be set individually (also contrast / brightness).
Pixel drop-outs:
- Manufacturer-related pixel drop-outs are always possible; up to 15 pixels (light or
dark) are permissible.
Backlight control:
- The backlight is automatically adjusted. This means that the maximum contrast that
is set is retained as long as the control is not set to stop.
- When set to control stop, the maximum contrast is will be less with lower emission or
luminance. For the minimum values, see the chapter "System-related Adjustments".
If there is a drop below the "minimum value", replace the display.

Repair / Troubleshooting

In a malfunction situation, the display is only replaced completely.
To adapt to the particular system, see chapter "Application-related Adjustments".
NOTE

Test and Auxiliary Equipment

All required settings, see the Adjustment chapter, must be performed using the
push-buttons on the side of the unit. A laptop is not required.
"SMFit test meter" luminous density meter, see the SPC.
Siemens AG
Medical Solutions
Every defective display must be returned with an exact description
of the error. It is virtually impossible to find sporadic and / or tem-
perature-related errors or also system-related causes without a
description of the malfunction.
TD00-000.841.02.04.02
05.05
Page 5 of 22
CSPS21
5
0
0
Display

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

30 80 991

Table of Contents