2.3 Inspection for Appropriate Connection

1) Check that the system is correctly operating. Refer to the
instruction manual for the Main unit.
2) Fill sterile water into a bucket and confirm that a biopsy guide
line and an echo image of test needle are overlapped correctly at
all selectable angles. (See Fig.5)
3) Relationship between the needle insertion groove and needle guide
line angle following "3. Operation Procedure" Fig.17. Also, Confirm
that the needle moves smoothly and there is no blurring between the
needle and biopsy attachment.
Sterile water
Biopsy guide
line (A dotted
line)
Before use, make sure there is no damage to the surface to which the
needle cannula is attached, and the biopsy attachment works properly.
Fig.5 Echo image of test needle
Test needle
Right-left
orientation mark
Needle guide angle
Needle echo
(A solid line)
CAUTION