1.3 Intended Use

The Biopsy Probe EUP-B715 is designed for observation and diagnosis
mainly of the following regions by connecting with the HITACHI
ultrasound scanner.
Biopsy (with Biopsy Attachment)
General abdominal organs

1.4 Components

The probe components of the EUP-B715 are as follows:
1) Biopsy Probe EUP-B715 ······························· 1 piece
2) Biopsy attachment for 12G-13G ······················· 10 pieces
3) Instruction Manual ·································· 1 copy
1) The biopsy attachment (Needle guide, Needle insert) is disposable.
Do not reuse. When the biopsy attachment is reused, it may run the
risk of exposing the patient to infectious diseases.
2) When using the needle cannula of the electrosurgical unit while using
the biopsy attachment as a guide, be careful not to damage the
insulation coating of the needle cannula. [When inserting or
removing the needle cannula into or from the biopsy attachment, you
may damage the insulation coating of the needle cannula, which may
cause a burn to tissue contacting the exposed section of the
insulation coating.]
1) Use appropriate biopsy attachment that is correspondent with the
needle that you use.
2) If you use Cool-tip 17G or LeVeen SuperSlim needle 17G of RFA
needle, use the biopsy attachment EZU-PA7B1-C surely.
3) Sterilization has not been made to the probe and biopsy attachment
shipped from the factory.
Prior to use of them, be sure to clean and sterilize them.
4) Do not hold the needle cannula of the electrosurgical unit with metal
tweezers, forceps, and the like. [Doing so may damage the insulation
section of the needle cannula, and may cause a burn to a non-treated
area.]
5) A biopsy should be performed only by a well-trained physician.
for 14G-16G ······················· 10 pieces
for 17G-19G ······················· 10 pieces
for 20G-23G ······················· 10 pieces
for cool-tipTM (17G), LeVeenTM
SuperSlim Needle(17G) ············· 10 pieces
WARNING
CAUTION
-2-
Q1E-EP1245