5.1

Point of use (Pre-cleaning)

Pre-cleaning should be done immediately after each
use. The procedure is as follows:
A) C41L47RP probe
1) Remove the protective cover.
2) Clean the probe of all patient's blood or fluid with running tap
water until the surface of the probe looks visually clean.
3) Wipe the whole surface of the probe with gauze pad and remove
superficial visible impurities.
B) Magnetic sensor attachment
1) Remove the Magnetic sensor attachment and the magnetic sensor from
the probe.
2) Immerse the Magnetic sensor attachment in sufficient amount of high
quality tap water. Scrub it using soft cloth to remove all visible
soil from its surface.
5.2

Containment and transportation

Putting the contaminated equipment into exclusive
shock and damage proof container for transportation is recommended.
It is recommended that instruments are reprocessed as soon as possible
and not later than 4 hours after usage.
- 18 -
Point of use
(Pre-cleaning)
Containment and
transportation