7. Maintenance and Safety Inspection

7.1

Daily Inspection

7.1.1
Visually inspect the surface of the probe head, housing,
cable and connector for any crack, scratch or
denaturalization. If you find any damage, do not use the
probe and contact a service support immediately.
7.1.2
Visually inspect the Puncture Guide Fixture for deformation
or denaturalization. If you find any damage, do not use
the Puncture Guide Fixture and contact your local HITACHI
distributor immediately.
7.1.3
Visually inspect the surface of the Magnetic sensor
attachment for any crack, deformation or denaturalization.
If you find any damage, do not use it and contact a service
support immediately.
7.2

Store

After using the probe and accessory, they should be cleaned
and disinfected/sterilized according to "5. Cleaning, Disinfection
and Sterilization" and "6.Cleaning, Disinfection, and Sterilization of
EZU-PA3U" immediately. Then store them in a cool, dustproof
and dark place avoid high temperature and humidity direct
sunlight.