Manual disinfection:
1) Prepare the disinfectant solution as stated in the procedure for
the probe.
2) Immerse the parts of the EZU-PA3U into the disinfectant. Set a clock
to insure the recommended contact time which is 5 minutes.
3) Rinse the parts of the EZU-PA3U with deionized water for 1 minute.
(alternatively: immerse them in a tray filled with deionized water
for 5 min.)
4) Visually check the outer surface of the parts of the EZU-PA3U for
leavings of the disinfectant. If necessary, repeat the rinsing.
6.4

Automated cleaning and disinfection

The following items must be provided prior
to automated cleaning and disinfection:
a) Washer disinfector: according to DIN EN ISO 15883 with chemo-thermal
program (temperature: max 60° C).
b) Detergent: Korsolex Endo-Cleaner (Bode Chemie; # 972 020) or another
cleaning agent with approved material compatibility for this medical
device
c) Disinfectant: Korsolex Endo-Disinfectant (Bode Chemie; # 972 030)
or another disinfectant with approved material compatibility for
this medical device
Washer Disinfector
Automated Cleaning
Automated Disinfection