B) Magnetic sensor attachment
1) Prepare the disinfectant solution as stated in the procedure for
the probe.
2) Immerse the Magnetic sensor attachment into the disinfectant (see
Fig. 10). Set a clock to insure the recommended contact time which
is 5 minutes.
3) Rinse the Magnetic sensor attachment with deionized water for 1
minute. (alternatively: immerse it in a tray filled with deionized
water (see Fig.10) for 5 min.)
4) Visually check its outer surface for leavings of the disinfectant.
If necessary, repeat the rinsing.
Water, Detergent
or Disinfectant
Fig.10
Immersion of the probe and the Magnetic
sensor attachment.
5.4

Drying

1) Wipe the equipment with a single-use, fluff-free wipe or towel to
remove moisture from the surface of the equipment.
2) Dry the equipment naturally in an ambient temperature between 15-30°C
for a minimum of 4 hours. Alternatively the equipment can be dried
using a drying heater at a temperature of less than 60°C.
- 22 -
Drying
Q1E-EP1451