Detector has been tested with below table' s X-ray condition. This table is only for reference. The legally certified
radiologist expert should control X-ray dose.
SID (inch / cm)
Tube Voltage (KV)
Chest P-A
72 inch / 182.8 cm
120 KV
C-spine LAT
72 inch / 182.8 cm
70 KV
L-spine A-P
40 inch / 101.6 cm
75 KV
Abdomen A-P
40 inch / 101.6 cm
75 KV
Pelvic A-P
40 inch / 101.6 cm
70 KV
Wrist A-P
40 inch / 101.6 cm
55 KV
Elbow A-P
40 inch / 101.6 cm
55 KV
Shoulder AP
40 inch / 101.6 cm
70 KV
Foot A-P
40 inch / 101.6 cm
58 KV
Ankle A-P
40 inch / 101.6 cm
59 KV
Knee A-P
40 inch / 101.6 cm
58 KV
Regarding paediatric dosage, it should be much less than for an adult. The certified radiologist should pay
special attention to paediatric x-ray dosage levels.
Adult
Tube Current (mA) Tube Current x Time
(mAs)
250 mA
2.5 mAs
200 mA
10 mAs
320 mA
20 mAs
320 mA
10 mAs
320 mA
16 mAs
100 mA
2.5 mAs
100 mA
3.2 mAs
200 mA
6.4 mAs
100 mA
2.5 mAs
100 mA
3.2 mAs
100 mA
6.4 mAs
GRID
Item
SID
100 cm / 130 cm / 150 cm / 180 cm
Size
460 x 460 mm
Ratio
10 : 1
Frequency
215 Line / Inch
Inter Spacer
AL
Recommended Specification