INSTRUCTION MANUAL 10DX3111A&G Sizes 14 - 24 Inches

1.3 Specifications

Power Requirements
Power Consumption
Flowmeter Characteristics:
Meter Size/Flow Capacity
Span
Rangeability
Minimum Fluid Conductivity
System Accuracy:
Frequency Output
4-20 mA Output
Meter Capacity
Environmental Limits:
Enclosure Classification
(Standard Meter)
Accidental Submergence Option IEC 529 IP67; NEMA 4X, 33 ft. H
Continuous Submergence
Option
Relative Humidity
Ambient Temperature Limits
at primary
1-6
See 1.2 Model Number Breakdown
23VA typical
Refer to Converter Instruction Bulletin, as specified,
to supplement this data.
See TABLE 1-2.
Factory set at specified range between extremes listed in
TABLE 1-2; can be field adjusted.
100:1
5 µS/cm
Flowrate
<2% of meter capacity
>2% of meter capacity
specified on primary data tag (equal to max flow capacity
in engineering units). This factor is equivalent to a flow ve-
locity of 10 m/s in the metering tube.
IEC 529 IP65; NEMA 4X, indoor or outdoor installation
IEC 529 IP68;NEMA 4X, 33 ft. H
10% to 90% (non-condensing)
o
o
Max: 140
F(60
C) for Process Temperatures up to 266
o
(130
C). Max. temperature is reduced by 1
gree increase in process temperature above 266
o
o
Min: -13
F (-25
C)
Accuracy
+0.01% of meter capacity
+0.5% of rate
add +0.1% of span to above
numbers
O/48h (10m H
2
O depth (10m H
2
o
for each de-
O/48 h)
2
O depth)
2
o
F
o
o
F (130
C)