Commissioning

Enabling the Framegrabber Service Modes

Press the "OK/Enter" key on the control panel. The control panel displays the main
menu.
Press the "Erase/Cancel" key on the control panel 4 times.
Press the "Color Capture / Time" key on the control panel once. The display shows "Key
not supported...".
Press the "OK/Enter" key on the control panel.
Press the "Menu" key on the control panel.
Select the "Installation" menu line with the cursor key "
key.
NOTICE

Programming the external exposure release

To enable the framegrabber to evaluate the print key on the Memoskop and SIREMOBIL
basic unit and to print an image, this function must be programmed.
The "Installation" menu is already selected. See "Enabling the Framegrabber Service
Modes".
If not, select the "INSTALLATION" menu line with the cursor key "
"OK/Enter" key.
Select the "Exam ID Config" menu line with the cursor key "
"OK/Enter" key.
Select the "Auto Date/Time" menu line with the cursor key "
"OK/Enter" key.
Press the "Menu" key to select the main menu again.
Siemens AG
Medical Solutions
If the installation menu cannot be selected, press the OK/Enter
key again and follow the instructions above under "Enabling the
Framegrabber Service Modes". The installation menu remains
enabled and selectable until the framegrabber is switched off.