Automatic sizing with existing 50/60 Hz monitors
NOTICE
•
Select the"INSTALLATION" menu line with the cursor key "
Enter" key.
•
Select the "Select/Setup Host" menu line with the cursor key "
panel of the framegrabber and press the "OK/Enter" key.
•
Select the line "SIEMENS" with the cursor key "
•
Select the correct videofrequency and polarits (see above) with the cursor key "
and press the "OK / Enter" key. The actual programmed value is marked with "
beginning of the line.
•
Select the"Automatic Sizing" menu line with the cursor key "
Enter" key.
•
Confirm the "Automatic Sizing Proceed?" query with the "OK / Enter" key.
•
Press the "Menu" key to select the main menu again.
Programming the film type on the printer
•
Insert the film supplied in the film compartment of the printer.
Program the film type.
See operating instructions, section "Loading DirectVista Media".
SIREMOBIL Compact
System Manual
Carry out only with existing 28 44 376 X2183 monitors.
In the supply state of the framegrabber, the geometric dimensions
are optimized for monitor output signals of 100 or 120 Hz. With
existing 28 44 376 X2183 monitors with refresh rates of 50 Hz or
60 Hz, the text overlay is not fully displayed on the film.
For corrections, automatic sizing of the framegrabber must be
carried out in these cases.