an associated Help screen that will automatically display when moving
from screen-to-screen. Click on the Help or the Entry screen to bring it to
the front, as needed. The initial Help screen includes a Read Me First
area at the top of the screen. This should be read thoroughly.
As information is entered, the NGC's oven will warm up, and diagnostics
will run. The diagnostics cannot finish until the oven is at the correct
temperature and stabilized. The start-up procedure cannot be finished
until the diagnostics are complete. Depending on the ambient
temperature, this can take up to 30 minutes or longer.
During initial start-up, all streams are disabled. The last phase of the
diagnostics is Stream Tests, and streams that have input pressure are
re-enabled. Therefore, if a stream is going to be used, it should be
plumbed in and sample pressure applied so that the stream can be
tested during start-up; however, a stream can always be added and
enabled at a later date.
15A
Step through all the screens in the Start-Up Wizard, and fill in the
required information. Be sure to put the unit in Run mode, as
instructed, and allow the unit to run anywhere from eight (8)
hours to overnight.
Step 16
Calibrate the NGC
After the unit has run for a minimum of eight (8) hours, it will need to be
calibrated.
16A
Connect the MMI cable to the unit, and start PCCU. If using a
USB connection, a screen may display asking about a
partnership. Click Cancel, or close the screen.
16B
On the Operation screen, the unit should be in Run mode. Click
the Hold button. The Next Mode indicator will illuminate, and the
unit will go into Hold mode at the end of the cycle. The user can
go on to the next step without waiting for the end of the cycle.
16C
On the Operation screen, click on the Calibration button to
display the Calibration Setup dialog box. Within this area, the
user will need to verify the following calibration information:
• Verify that the calibration stream being used (First Calibration
Stream) is correct; change if needed.
• There are default values in the Calibration Cycles Average and
in the Purge Cycles windows. The user can change these
values now even if they changed them during the Start-Up
Wizard.
• Verify that the values in the % Blend 1 column match the
calibration blend bottle and Total Mole % is 100. If not 100%
9