3) Similarly, set the hour and minute in
turn.
4) Press the
button. (Timer operation starts.)
The sterilization temperature appears on
the digital display Ⅰ and the date and
time appear alternately on the digital
display Ⅱ .
To cancel the timer operation, press the
button.
Timer setting
Year, month, day, hour and minute have been set at the time of
shipment.
If the timer setting is needed anew after replacement of parts, etc.,
set it in the procedure as follows:
When the pushbuttons
year entry appears and flickers.Enter Christian year by
buttons and press the
Enter the month, day, hour and minute in the same manner.
When setting the minute, press the
time signal.

On Power Failure

In power failure, all of the control panel displays are cleared.
When power failure occurs or when the power switch has been set
to OFF, the safety device works to lock the cover, making it impossible
to open the cover.
To use the product after restoration of power
When the power is restored, the control panel displays are in standby status.
1) Check the following points before opening the cover.
Pressure is lowered to 0 MPa.
COVER LOCK lamp goes out.
When the COVER LOCK lamp is lit, the temperature or pressure
inside the chamber rises to make it impossible to pull the movable
handle toward you.
2) Take out the sterilization objects
If liquid boils over in liquid sterilization, clean the inside of the
chamber.
Inside of the chamber : See page 26.
3) Check the sterilization by sterilization indicator.
Objects can not be sterilized in some cases.
4) If objects can not be sterilized, repeat the operation.
and then
are pressed, the Christian
button.
button to set exact time by
and
Use care not to scald yourself when
taking out the sterilized objects at the
end of operation.
Be sure to wear thermally
insulated leather gloves. Note
that it takes time until liquid is
cooled down.
Handling a sterilized object with
bare bands immediately after the
end of operation may result in burns.
Lamp goes out.