8.Set example of parameter
  8-1.Flow of parameter set 
            Parameter should set according to the flow as shown hereunder. 
Measure 
window 
※ In case setting up value over input range exceeding specification, setup value  
of parameter will indicate 「Err」on indication L E D, let you know setup value  
is wrong and do not accept next operation. 
※ In this case, please change to correct set value again. 
      8-2.Example of parameter set 
          Following is a set example to set integral time at 15 second. 
          (Providing that set value was 0 second before change.) 
          (1)  Push  MOD button  1  time・・・・・・・・・・move  to  parameter  window  and  LED(orange). 
          (2)  Push SEL button 5 times・・・・・・・・ LED(orange)lit goes down in turn 
                                                           and INTEGRATION TIME LED will light. 
          (3)  By using both → button and ↑button, 
               set Indication LED at「15」. 
          (4)  Push MOD button1 time・・・・・・・・・return  to  measure  window  and  start  measure 
again at integral15 second. 
          You operate other parameters with same procedure, please. 
          In case setting other parameters continuously, at (4) above instead of MOD you push 
SEL and can perform both setting parameter and moving to plural items at the same time. 
          Reference: Parameter and light up of corresponding LED 
RANGE             :  only RANGE LED light up 
L1 ON            :  only L1 ON LED light up 
                  L1 OFF           :  only L1 OFF LED light up 
                  L2 ON            :  only L2 ON LED light up 
                  L2 OFF           :  only L2 OFF LED light up 
                  INTEGRATION TIME :  only INTEGRATION TIME LED light up 
9.Maintenance  
    Measure probe should regularly be wiped and cleaned with dry rag. 
You please keep it cleaned once in every 3 months but more often because of changing media 
or measure environment, you see. 
When cleaning, you please switch off power. 
Temperature of pipe may be high and dust monitor might be heated accordingly. 
Be careful! Or you might get your hands burned.  
Caution: The earth line should not be removed even at the maintenance in case 
!
               it may cause a fault. 
Parameter mode 
MOD button 
Press down 
(to    parameter 
mode) 
Push down MOD button  
(after parameter set, to  measure 
window) 
Parameter select 
SEL button 
Push down 
Respective 
parameter select 
written Range on the left will light on. 
- 11 -
Value set 
→ Move  beam  by 
button 
↑ Each  button 
push down change 
value 
SEL button push down 
(after set parameter move 
to other parameter item) 
 