Download Print this page

Kit Installation - ABB 3KXA494400L005 Installation Procedure

Universal sensor accessory in open channel and open tank mounting kit

Advertisement

6
U N I V E R S A L S E N S O R A C C E S S O R Y ¾ I N | O P E N C H A N N E L A N D O P E N TA N K M O U N T I N G K I T | I N/A N A I N S T/0 4 9 - E N R E V. A

4 Kit installation

Referring to Figure 3:
1 Pass M3 screw
(supplied) through the end of sensor chain
A
mounting adaptor
B
, feed it through the last link on
opposite end of chain
C
hole in opposite side of sensor chain mounting adaptor
Lock into position using supplied lock nut
2 Pass EZLink
sensor cable connector
through the hole in sensor chain mounting adaptor
feed entire cable
F
and chain
3 Screw sensor body
G
onto sensor chain mounting
adaptor
B
using PTFE tape to seal.
NOTICE
Do not overtighten.
4 When the required chain/cable length has been determined,
secure the chain onto second threaded stud using locknut
(supplied).
NOTICE
Ideally, the measuring end of the sensor should be a
minimum of 300 mm (12 in) below the free surface of the
liquid being monitored.
5 Gently lower the sensor into the sample until the sensor is
fully immersed to the required depth.
6 Connect the EZLink sensor cable connector to the AWT420
transmitter.
WARNING
Ensure the cable/chain run is routed securely and safely and
does not present a trip hazard.
7 Secure the chain to the installation. Make sure that it does
not come in contact with any wall or solid object.
and screw M3 screw
A
into tapped
D
.
and sensor cable
E
C
length through.
B
.
F
B
and

Advertisement

loading

This manual is also suitable for:

3kxa494400l006