6 - 2
The oil temperature in the single tank is measured. If the temperature is not within the
range +8°C to +50°C, an error will be displayed.
The positions of service switches "S1" and "S2" on the CPU board D1 are queried.
PC mode is used only for testing procedures at the factory!
The exposure parameter display indicates "PC-Mode" and the selection buttons on the
Control and display panel are disabled!
The charging procedure for the capacitor bank is completed after reaching the final
voltage value. According to the residual voltage in the capacitor bank, this takes between
5 and 120 s.
The green "ready" LED on the Control and display panel is turned on.
The unit is ready to operate!
MOBILETT Plus/ Plus E
The switch positions are:
"S1" and "S2"
"S1"
"S2"
Register 6
SPR8-215.850.02
Rev. 01
Power-up sequences
off
=
normal mode
on
=
software-supported service mode
on
=
PC mode
Page 2 of 2
01.02
SPS-UD
Siemens-Elema AB