12 - 2
Exposure release
•
the yellow exposure-indicator lamp on the Control and display panel is turned on,
•
the high-voltage inverter is switched on,
•
the resulting tube current is integrated until the preset mAs value is reached, followed by;
•
the high-voltage inverter is switched off,
•
preheating is resumed
•
a long beep tone is generated.
•
the exposure-indicator lamp is turned off.
•
the rotating anode is braked.
NOTICE
MOBILETT Plus/ Plus E
The secondary contact in the exposure-release switch is closed. This initiates the
following additional functions:
If the MOBILETT Plus is equiped withe a remote
exposure switch system, please see Installation
Instruction: Installation of remote exposure switch
system, chapter 3 "Supplement to the Instructions for
Use".
Register 6
Preparation and exposure release
SPR8-215.850.02
Rev. 01
01.02
Page 2 of 2
Siemens-Elema AB
SPS-UD
12
Solna, Sweden