Temperature Readings
Displays internal temperatures for the Power Board, Therapy Board, Printer, and Battery. These readings may vary
according to fan speed or external conditions. An abnormally high reading may indicate a fan not working or other fault.
The internal fan should turn on at the following temperatures:
Fan ON (High)
65º
Power Board
65º
Therapy Board
N/A
Printer
N/A
Battery
Fan Test
Select Off, Low, or High
Note: The Low fan speed may be difficult to hear in a noisy environment.
2025653-048 Revision B
Section 5: Performance Verification and Safety Testing
Fan ON (Low)
Fan OFF
60º
55º
60º
55º
N/A
N/A
N/A
N/A
™
Responder
2000