Temperature Readings; Fan Test - GE Responder 2000 Service Manual

Defibrilator/monitor
Hide thumbs Also See for Responder 2000:
Table of Contents

Advertisement

Temperature Readings

Displays internal temperatures for the Power Board, Therapy Board, Printer, and Battery. These readings may vary
according to fan speed or external conditions. An abnormally high reading may indicate a fan not working or other fault.
The internal fan should turn on at the following temperatures:
Fan ON (High)
65º
Power Board
65º
Therapy Board
N/A
Printer
N/A
Battery

Fan Test

Select Off, Low, or High
Note: The Low fan speed may be difficult to hear in a noisy environment.
2025653-048 Revision B
Section 5: Performance Verification and Safety Testing
Fan ON (Low)
Fan OFF
60º
55º
60º
55º
N/A
N/A
N/A
N/A
Responder
2000
Page 57

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents