9.
5
4
K1
9
Fig. 6
10.
11.
12.
13.
14.
15.
16.
Mobilett Plus/ E/ M/ HP
Service
Feed the free end of the cable with the module connector from around the back
and into the right side of the machine.
7
D1
Mount the electronics package (pos 1 / Fig. 6) alongside the main Mobilett CPU
board with the double-sided tape. Make sure that the distance between the right
side metal wall (pos 2 / Fig. 6) and the electronics package is at least 15mm. Plug
the module connector from the remote sensor into the package (pos 3 / Fig. 6)
Unplug the existing connector from D1K1.
Using the IDC connector supplied, route the PCB flat cable behind the control
panel and connect it to D1K1 (pos 4 / Fig. 6).
Fold the flat cable and fasten the ferrite core around it (pos 9 / Fig. 6) as close to
the K1 connector as possible.
Reconnect the unplugged connector to the extra IDC connector.
Fasten the ground connector cable (pos 10 / Fig. 6). Make sure the washers are
oriented as described in Fig. 6.
Position the provided tie-mounts to the areas indicated in Fig. 6 at pos 5, 6 and 7.
2
6
8
1
10
MOB00665
SPR8-220.031.02
Rev. 02
11.02
Installation
washer
circuit board
3
aluminum cover
Page 4 of 8
Siemens-Elema AB
SPS-UD
washer
Solna, Sweden