CHAPTER 3 3 Environmental
CHAPTER
5322 871 63303
96.00
Environmental
Environmental Conditions
Environmental Conditions
3.1 Operating Conditions
Ambient temperature range.
operation within specification
all functions operational
Max. permissible rate of temperature change 1°C/hour, 0.1°C/min.
Relative humidity at 20°C
Ambient air pressure
Dust class (average industrial environment) 100,000
3.2 Static Electricity
If the factory in which the system will be installed is ESD (ElectroStatic
Discharge) safe, no special measures are required for operator use.
However, special attention must be paid to electrostatic discharge prevention
during maintenance activities. These measures must be known and applied by
all maintenance engineers.
Pre-installation Manual CSM 66/84/84V III
PA 1306/40 /50 /60 /61
Conditions
Conditions
from +20 to +28 °C
from +15 to +35 °C
80%, dew point below 18 °C
from 800 to 1100 mbar
11