Mains Voltage Supply And Earthing; Specification Of Mains Voltage Supply; Specification Of Earthing; Local Area Network (Lan) - Philips CSM 66 III Preinstallation Manual

Table of Contents

Advertisement

5322 871 63303
96.00
Services Required
Services

4.2 Mains Voltage Supply and Earthing

The system can be connected to most common single-phase mains supplies,
from 200 V to 240 V, 2-wire, with one earth wire. For other mains supply
configurations, the customer must take measures to adapt the local mains
supply configuration to the required configuration.

4.2.1 Specification of Mains Voltage Supply

Mains supply required
Power consumption
Power factor
Frequency
Frequency stability
External fusing
Line voltage fluctuation
Connections
Cross-section of mains cable

4.2.2 Specification of Earthing

Connection:
Earth resistance: Less than <0.1 ohm.

4.3 Local Area Network (LAN)

Not applicable for this machine.

4.4 External Emergency Circuit

Not applicable for this machine.
Pre-installation Manual CSM 66/84/84V III
PA 1306/40 /50 /60 /61
Required
200-240 Volts ± 10% Single phase
2.0 kVA (nominal)
4.0 kVA (maximum)
0.85
50 Hz or 60 Hz
±2%
16 A slow-blow
+ 10% (slow variations and short during
transients
via 3 wire cable (see fig. 4.1)
(cable diameter 1.8 mm),
3 x 2.5 mm
The system must be connected to factory earthing via the
mains cable.
2
15

Advertisement

Table of Contents
loading

Table of Contents