Quick Start Guide

9.4.2 Focus tube

When to use the focus tube:
• Short distance from line / component of 0.05 m (0.16 ft)
• Freely accessible line / component
• The lines and components being inspected are physically very close to each other
• Medium to high noise
• Use if there is no gooseneck available

9.4.3 Gooseneck

The gooseneck should be used when the lines and components being inspected are physically very
close to each other. In addition, the gooseneck's shape can be adjusted as necessary in order to
make it possible to comfortably inspect lines and components that are difficult to access.
The gooseneck's sensitivity has been reduced in order to attenuate noise. Accordingly, it is
exceptionally well-suited to inspecting compressed air components in a targeted and local manner
when there is high noise, such as is the case inside compressed air distributor cabinets and with
systems that use pneumatic cylinders.
The tip of the focus tube only lets very few ultrasonic waves
through in the direction of the ultrasonic transducer, making
it possible to pinpoint leaks very exactly.
Accordingly, it is recommended to use the focus tube at short
distances in order to accurately pinpoint the corresponding
leak.
Quantification distance: (Distance) -> 0 ... 0.2 m (0 ... 0.66 ft)
If you ordered the gooseneck separately, you will first need 
to import the corresponding data (which is found on the
included USB drive) into the METPOINT
able to use the gooseneck.
Import:
Home -> Export / Import -> Import -> New Sensor ->
Gooseneck xxxxxxxx
METPOINT
CID
®
CID in order to be
®
19 | 28