5.4.2 Cable Connection
 
LF622F
Connect and install the terminal-treated cables to the terminal block.
*Connect the cables to the terminal block securely. A loose connection may cause incorrect
measurement. After connecting a cable, try to pull it to check whether it has been
connected securely.
(1)Referring to combined converter's manuals of "Connections and Grounding", connect each
cable to the terminal block. Tighten the screws of the terminal block tightly to ensure the
secure connection. A loose connection may cause incorrect measurement. After connecting a
cable, try to pull it to see whether it has been connected securely. 
                       
             
(2) Attach the terminal box cover . To keep the seal, tighten securely the cover.
Figure 5.6 Connecting a Cable to Terminal Block
30
 
Phillips screwdriver
Terminal block
     
6F8A0943
* The appropriate torque for tightening
the terminal board screw is 1.2 N.m.