Siemens MOBILETT Plus HP Maintenance Instruction page 36

Hide thumbs Also See for MOBILETT Plus HP:
Table of Contents

Advertisement

36
2.2.10.3
Single tank fork
Fig. 15:
Pos. 1
Check without torque
SIM
Screw connections on the fork
Check and tighten all screw connections of the fork on the support arm
SIM
Single tank screw connections
Check the four screw connections of the single tank. They are accessible from the top
side of the single tank without removing the cover
2.2.10.4
Lubrication
Fig. 17:
Remove the cover of the column stand.
MOBILETT Plus HP
Fig. 18:
Pos. 11
SPR8-220.101.04.03.02
07.05
Inspection and Maintenance
Fig. 16:
(Fig. 16 / p.
Rotation wear
Page 36 of 64
CS PS 24
(Fig. 15 / p.
36).
36).
Siemens AG
Medical Solutions

Advertisement

Table of Contents
loading

Table of Contents