Inspection and Maintenance
2.2.9
Mains cable
Fig. 11:
Pos. 1
PMP
Replace the mains cable every 2 years.
The mains cable must be replaced every two years.
SIM
Damage
•
Pull the entire (6 m) mains cable out from the cable winder.
•
Check the mains cable and plug for damage. If necessary, replace them. See section
"Replacement of mains cable / cable winder" in the Service Instructions.
PMP
Cleaning
•
Clean the mains cable using a paper towel dampened with a lukewarm solution of wa-
ter and a household cleaning agent.
•
Pull the brake handle for the cable upwards.
SIM
Cable winder
•
Move the cable up and down while rewinding it, in order to avoid any slack, until the ca-
ble is completely rolled up. Check the proper functioning of the cable winder and brake.
SIM
Protective conductor
•
Disconnect the power supply plug.
•
Check the protective conductor connection on the plug. The protective conductor must
be securely attached and must be slightly longer than the other two wires. This applies
for all plug types.
•
Close the plug after inspecting it.