Inspection and Maintenance
PMP
Glide bushing on the connecting rod
Lubricate the glide bushing on the connecting rod
2.2.11

Arm cable harness

SIM
Damage
Check all the cables for damage. Closely inspect the areas that are frequently subject-
ed to torsion strain
Replace any damaged or broken cables. If you cannot determine any direct cause for
the damage, replace all cable harnesses. Refer to the section entitled "Replacement of
the support arm cable harness" in the Service Instructions.
2.2.12

Turn plate

Fig. 19:
Connect the batteries and switch the system to "B".
Move the unit to a level surface.
PMF
Button illumination
Press one of the turn plate buttons. The buttons must light up when the turn plate is acti-
vated.
PMF
Function
Check whether the system can be rotated around the turn plate
Press the turn plate button again and check whether the turn plate retracts to the park-
ing position.
Repeat the same test with the other turn plate button. If a problem arises, proceed ac-
cording to the section "Adjustment and replacement of the turn plate" in the Service In-
structions.