Proper use
This fully automatic electronic blood pressure monitor is intended for measuring the blood pressure at home. It is a non-invasive blood pressure measurement system to measure the diastolic and systolic blood pressure and pulse of an adult using a non-invasive technique by means of a cuff, which needs to be fitted on the wrist.
Contraindications
Follow the instructions for use! Non-observance of these instructions can result in serious injury or damage to the device. | |
These warning notes must be observed to prevent any injury to the user. | |
These notes must be observed to prevent any damage to the device. | |
NOTE These notes give you useful additional information on the installation or operation. | |
Device classification: type BF applied part | |
Not to be used by very young children (0-3 years)! | |
Keep dry | |
LOT number | |
Manufacturer | |
Date of manufacture |
Please check first of all that the unit is complete. The following parts are included as standard:
Please contact your supplier if you notice any transport damage on unpacking the unit.
Read the instruction manual carefully before using this device, especially the safety instructions and keep the instruction manual for future use. Should you give this device to another person, it is vital that you also pass on these instructions for use.
SAFETY NOTES FOR BATTERIES
Blood pressure is the pressure created in the blood vessels by each heart beat. When the heart contracts (= systole) and pumps blood into the arteries, it brings about a rise in pressure. The highest value is known as the systolic pressure and is the first value to be measured. When the cardiac muscle relaxes to take in fresh blood, the pressure in the arteries also falls. Once the arteries are relaxed, a second reading is taken – the diastolic pressure.
Please ensure that the polythene packing is kept away from the reach of children! Risk of suffocation!
systolic mmHg | diastolic mmHg | Bloodpressure indicator | |
≥ 180 | ≥ 110 | severe hypertension | red |
160 - 179 | 100 - 109 | moderate hypertension | orange |
140 - 159 | 90 - 99 | mild hypertension | yellow |
130 - 139 | 85 - 89 | high-normal blood pressure | green |
120 - 129 | 80 - 84 | normal blood pressure | green |
< 120 | < 80 | optimal blood pressure | green |
Blood pressure that is too low represents just as great a health risk as blood pressure that is too high! Fits of dizziness may lead to dangerous situations arising (e.g. on stairs or in traffic)!
You must insert the batteries provided before you can use your unit. The lid of the battery compartment is located on the left hand side of the unit. Open it and insert the 2 AAA LR03 type 1.5 V batteries supplied. Ensure correct polarity when inserting the batteries (as marked inside the battery compartment).
Close the battery compartment. Replace the batteries immediately, if the change battery symbol appears on the display or if nothing appears on the display after the unit has been switched on.
After insertion of the batteries you need to set the date and clock. To enter the setting mode, press and hold the - button for approx. 3 seconds. Firstliy, the user memory 1 or 2 is flashing. Now press the - button again to adjust the correct year figure. Adjust to correct year by pressing - button and press - button to confirm the setting. The device now switches to the month setting mode. Proceed accordingly (in the same way as with the year adjustment) to adjust the correct figures for month, day, hour and minute. Afterwards, the setting procedure is complete. Shortly, the current date appears and then the clock appears. If you want to look at the date later on, press the - button shortly. If the batteries are changed, the set information is lost and needs to be entered again.
The medisana wrist blood pressure monitor BW 315 gives you the opportunity to assign readings to one of two user memories, each of which provides 120 memory slots. Press - button to select the user memory or . Your selection will be saved by the device and will be considered in next blood pressure measurement respectively when the measurement result is saved. Your selection is valid until you make another selection.
After the cuff has been appropriately positioned, the measurement can begin:
If it is necessary to interrupt a blood pressure measurement for any reason (e.g. the patient feels unwell) the - button can be pressed at any time. The device immediately decrease the cuff pressure automatically.
This unit features 2 separate memories, each with a capacity of 120 memory slots. Results are automatically stored in the memory selected by the user. Press the - button to select the user ( or). Now press the - button and the average values of the last 3 measurements appear on the display. Press the - button again and the latest stored measurement value will be shown, together with the memory symbol „MR" and the memory location place „1" . Repeatedly pressing the - button displays the respective values measured previously (MR2, MR3 and so on). If you have reached the final entry and do not press a button, the unit automatically switches itself off after approximately 3 min. Memory recall mode can be exited at any time by pressing the - button . The oldest value is deleted, if 120 measured values have been stored in the memory and a new value is saved - at the same time, „Ful" shortly appears on the display.
If you are sure that you want to permanently remove all stored memories of a user, press and hold the - button for approx. 5 seconds, until CL appears in the display. Release the button. Now press the - button again to confirm the deletion.
The following error codes will appear on the display when a measurement is abnormal:
Display | Cause |
ERR1 | No pulse was detected. |
ERR2 | Movement or talking during measurement. |
ERR3 | Inflation process lasts too long. The cuff probably has not been positioned properly. |
ERR5 | Systolic and diastolic pressure are too far apart from each other. Repeat the measurement. If you constantly receive unusual measurement results, please contact your doctor. |
Problem | Cause and solutions |
No display although the device is switched on and batteries are inserted | Maybe the batteries are weak. Insert new batteries and pay attention to correct polarity. Check, if the device is working well. In the case of doubt, contact the service centre. |
Measurement can not be proceeded or measurement failed | Put on the cuff correctly. Repeat the measurement after relaxing 30 minutes. Do not speak or move during the measurement. |
Each measurement result strongly deviates from the one before | Repeat the measurement in correct way after relaxing 30 minutes. Pay attention to all the instructions in this manual and to the common factors of wrong measurements. Slightly deviating measurement results are quite normal, as the blood pressure is constantly changing. |
The measured values are remarkably different from values measured in a doctors surgery | Store the measured results and consult your doctor. It is not unusual to measure strongly deviating values outside of your home. |
Please contact the manufacturer resp. the service centre if you can't solve the problem. Do not disassemble the unit by yourself.
Remove the batteries before cleaning the unit. Clean the unit and the cuff using a soft cloth lightly moistened with a mild soapy solution. Never use abrasive cleaning agents, alcohol, naphtha, thinner or gasoline etc.. Never immerse the unit or any component in water. Be cautiious not to get any moisture in the main unit. Do not use the unit again until it is completely dry. Only inflate the cuff when it is in position around the wrist.
Do not expose the unit to direct sunlight; protect it against dirt and moisture. Do not subject the unit to extreme hot or cold temperature. Keep the unit in the storage box when not in use. Store the unit in a clean and dry location.
This product must not be disposed of together with domestic waste. All users are obliged to hand in all electrical or electronic devices, regardless of whether or not they contain toxic substances, at a municipal or commercial collection point so that they can be disposed of in an environmentally acceptable manner. Remove the batteries before disposing of the device/unit. Do not dispose of old batteries with your household waste, but at a battery collection station at a recycling site or in a shop. Consult your local authority or your supplier for information about disposal.
This blood pressure monitor meets the requirements of the EU standard for non-invasive blood pressure monitors. It is certified in accordance with EC Guidelines and carries the CE symbol (conformity symbol) "CE 0297". The blood pressure monitor corresponds to European standards EN 60601-1, EN 606011-2, EN 60601-1-11, EN 80601-2, EN 1060-1, EN 1060-3 and EN 1060-4. The specifications of EU Guideline "93/42/EEC of the Council Directive dated 14 June 1993 concerning medical devices" are met.
Electromagnetic compatibility: (see attached leaflet)
Name and model | medisana Blood pressure monitor BW 315 |
Display system | Digital display |
Memory slots | 2 x 120 measurement values |
Measuring method | Oscillometric |
Power supply | 3 V, 2 x 1.5V batteries AAA LR03 |
Blood pressure measuring range | 30 – 280 mmHg |
Pulse measuring range | 40 – 200 beats/min. |
Maximum error tolerance for static pressure | ± 3 mmHg |
Maximum error tolerance for pulse rate | ± 5% of the value |
Pressure generation | Automatic with pump |
Deflation | Automatic |
Automatic switch-off | After approx. 3 minutes |
Operating conditions | +5°C to +40°C, 15 to 85% max. relative humidity |
Storage conditions | -10°C to +55°C, 10 to 95% max. relative humidity |
Dimensions (main unit) | approx. 63 x 63 x 29 mm (L x W x H) |
Cuff | 12.5 - 21.5 cm for adults |
Weight | approx. 110 g without batteries |
Artikel-Number | 51072 |
EAN-number | 40 15588 51072 4 |
In accordance with our policy of continual product improvement, we reserve the right to make technical and visual changes without notice.
The current version of this instruction manual can be found under www.medisana.com
Please contact your supplier or the service centre in case of a claim under the warranty. If you have to return the unit, please enclose a copy of your receipt and state what the defect is.
The following warranty terms apply:
medisana GmbH, Carl-Schurz-Str. 2,
41460 NEUSS, GERMANY
The service centre address is shown on the attached leaflet.