Error messages and instructions
88

Technical faults without fault number

Problem
The machine display
is dark.
The wash water in
the wash chamber
of the machine is
not drained after a
programme cancel-
lation.
Cause and remedy
The machine display is not showing any in-
formation. This can be as a result of the fol-
lowing:
– The display is faulty
– The connecting cable between the display
and the controls is faulty
– The plug connection is faulty
Call the Customer Service Department.
The machine is not switched on.
Press the I power switch.
An external residual current device (RCD)
has switched.
Inform building services.
An external emergency off switch was activ-
ated.
Inform building services.
Drainage after disinfection
The
enabled in the programme header of the re-
processing programme.
Manually disinfect the wash water and re-
move it from the wash chamber with the
Manual disinfection
ation, programme cancellation, pro-
gramme cancellation with manual disinfec-
tion").
parameter is
programme (see "Oper-