Manual Cleaning And Disinfection; Manual Cleaning - Hitachi ASU-1014 Instruction Manual

Electronic convex sector scanner
Table of Contents

Advertisement

MN1-5933 Rev.5

5-5. Manual cleaning and disinfection

Prepare following items before manual cleaning and disinfection.
Scanner
1) Detergent: ENZOL
approved material compatibility for this medical device.
2) Disinfectant: Cidex
material compatibility for this medical device.
3) 2 tanks, 1 for cleaning and 1 for disinfection - optional: 1 additional tank for rinsing with
deionized/tap water. (sufficient size for immersion of the submergible part of the scanner at full
length)
4) Soft, fluff free cloth or single use towel
5) Personal protective equipment (gloves, water repellent protective skirt, face protection mask or
protective glasses see also instructions of the manufacturer for the detergent and the disinfectant)

5-5-1. Manual cleaning

Scanner
1) The temperature of the detergent solution should be between 15-30 °C [59-86 °F], concentration is
1.6%.
Please note the minimum contact time of the detergent in the manufacturer's instruction.
If a differing detergent is used, please also consider the approved material compatibility for this
scanner.
2) Immerge the submergible part of the scanner (see figure) without connector into the detergent.
3) Wipe the submergible part of the scanner under the surface of the detergent solution with a single-
use, fluff free soft cloth to remove all visible soil. Be sure that all grooves of the scanner are
implemented during the cleaning process. If necessary use an appropriate cleaning brush for this
purpose.
4) Wipe the non-submergible parts of the scanner with a soft cloth dipped with a detergent.
5) Rinse the submergible part of the scanner with running tap water for 1 minute.
6) Alternatively to step 5 suspend the submergible part of the scanner in a tray filled with deionized
water/tap water for 5 min.
7) Visually check the outer surface of the scanner for cleanness. If necessary, use magnifying glass for
visually check. If there is still soil visible, repeat all above steps.
/Cidezyme
(Johnson & Johnson, #2258) or another cleaning agent with
®
®
OPA (Johnson & Johnson, # 20391) or another disinfectant with approved
®
-24-

Advertisement

Table of Contents
loading

Table of Contents