Sterilization; Liquid Sterilization (Usa Only); Storage - Hitachi ASU-1014 Instruction Manual

Electronic convex sector scanner
Table of Contents

Advertisement

MN1-5933 Rev.5

5-11. Sterilization

See "Table 1. Applicable cleaning
Follow the instructions of the sterilizer manufacturer regarding usage, temperature and sterilization-time
etc. Handling and maximum input to chamber of sterilizer should be according to operation manual of the
sterilizer.

5-11-1. Liquid sterilization (USA only)

• Applicable chemical solution for sterilization
The applicable sterilants are listed below.
General name
Hydrogen peroxide
Peracetic acid
Glutaraldehyde
Glutaraldehyde
Glutaral
Note: * indicates that the marked sterilant is not applicable in Canada.
Warning
After chemical sterilization, thoroughly rinse the scanner with sterile water.
Residual sterilant can cause an adverse reaction to the operator or patient.

5-12. Storage

For details about the storage environmental conditions, see section 2-5-2 "Storage environmental
conditions".
disinfection and sterilization methods" for available sterilization methods
,
Trade name
*
PERASAFE
®
Practical liquid 1.62W/V%
Acecide
*
®
Solution 6%
WAVICIDE
-01 *
®
Solution 2.65%
STERANIOS *
Solution 2.0%
Cidex plus
®
Solution 3.4%
-30-
Manufacturer
ANTEC INTERNATIONAL
Saraya Co., Ltd.
Medical Chemical Corporation
Laboratoires ANIOS
ADVANCED STERILIZATION PRODUCTS
A Johnson & Johnson company
Division of Ethicon, Inc.
®

Advertisement