7.2.2 Flow rate indicated is not correct
 
                                                              
START
               
 
 
 
 
Is the flow range correctly
 
set?
 
 
 
 
 
 
Is zero point correctly set?
 
 
 
 
 
Is the excitation current
 
 
value as stated on the
 
flow direction tag?
 
 
 
 
 
Is the inside wall of detector
 
pipe contaminated?
 
 
 
 
 
Is the output load resistance
 
 
as specified?
 
 
 
 
 
Are there two load
 
resistors connected to
 
the output in parallel?
 
 
 
 
 
 
Is accuracy calculated as follows?
 
(Measured flow rate)-(Actual flow rate)
 
Actual flow rate
 
 
 
 
 
 
Contact your nearest Toshiba
 
 
representative.
NO
YES
NO
YES
NO
YES
YES
NO
NO
YES
YES
NO
×
 
100%
YES
 
Set correctly.
Refer to combined converter's manual.
Perform the zero adjustment.
Refer to combined converter's manual.
Set correctly.
Refer to combined converter's manual.
Clean the inside of the detector pipe.
Design the output circuit so that the
output load resistance is less than
750
.
ohm
Connect the two load resistors in
series, if necessary.
NO
Calculate as shown on the left.