4.1 Notes on Selecting the Installation Location

1.
Avoid places within the immediate proximity of equipment producing electrical interference (such as
motors, transformers, radio transmitters, electrolytic cells, or other equipment causing
electromagnetic or electrostatic interference).
2.
Avoid places where excessive pipe vibration occurs.
3.
Avoid places where fluid runs in a pulsating form.
4.
Avoid places where there is direct sunlight. If this is unavoidable, use an appropriate shade
5.
Avoid places where corrosive atmospheres or high humidity conditions obtain.
6.
Avoid places of too great an elevation or constricted areas where clearance for installation or
maintenance work is not provided.
7.
Design piping so that the detector pipe is always filled with fluid, whether the fluid is flowingor not.
8.
The detector has no adjustable piping mechanism. Install an adjustable short pipe where needed.
9.
Chemical injections should be conducted on the downstream side of the flowmeter.
Separate
10. The standard length of the cable that connects the detector and converter is 30m. Select the
converter installation location so that the distance of the detector and converter will not exceed 30m.
 

4.2 Mounting Procedure

4.2.1 Pipe checks
(1)Before installing pipes, check for any leaning or misplacement (or eccentricity) as illustrated in Figure
4.1. An attempt to unreasonably connecting pipes that are inclined may lead to a detector breakdown or
fluid leakage. Connecting pipes in an eccentric state may also cause local wears and tears of linings and
grounding rings, as well as measurement errors.
Before installing pipes, make sure to flash the interior of the pipes to remove deposited matters. 
(a) Pipe leaning
Inclination
Figure 4.1 Pipe leaning and axis misplacement
−  15  − 
(b) Pipe axis misplacement (or eccentricity)
Eccentricity
6F8A0882 
 