HP 3457A Service Manual page 100

Hide thumbs Also See for 3457A:
Table of Contents

Advertisement

PERFORMANCE TEST CARD
90 DAY LIMITS
Hewlett-Packard Model 3457A
Digital Mutimeter
Serial Number
REAR INPUT TESTS (STANDARD
INSTRUMENT)
Test Performed by
Date
Reference Temperature
Step#
Input
to
Set-Up
and
High
Reading
Low
Test
Test
3457
Configuration
Limit
Limit
Pass
Fail
DC
VOLTAGE
OFFSET
TEST
(Rear
Terminals)
1
Short
300 V Range
= + 000.0007 v
- 000.0007 v
a
2
Short
30 V Range
+ 00.00020 v
= 00.00020 v
——
3
Short
3 V Range
+ 0.000007 v
- 0.000007 v
eS
4
Short
300 mV Range
+ 000.0040 mv
- 000.0040 my
5
Short
30 mV Range
= +: 0.00385 mv
- 0000385 my
DC CURRENT OFFSET TEST (Rear Terminals)
6
Open
300 UADCI Range
+ 000.0104 uA
- 000.0106 ua
7
Open
3 mADCI Range
+ :0.000104 ma
+ 0.000106 me
8
Open
30 mADCI Range
+ 00.00104 mA
+ 00.0010 me
9
Open
300 mADCI Range
+ 000.0204 mA
- 000.0204 me
10
Open
3 ADCI Range
+ 0.000604 A
- 0.000604 A
os
2-WIRE OHMS OFFSET TEST (Rear Terminals)
"1
Short
30 Ohm Range
+ 00.20335 ohm
= 00.20335 ohm
12
Short
300 Ohm Range
+ 000.2035 ohm
- 000.2035 ohm
13
Short
3 KOhm Range
— +: 0.000207 Kohm
- 0.000207 Kohm
4
Short
| 30 KOhm Range
+ 00.00027 Kohm
+ 00.00027 Kohm
15
Short
| 300 KOhm Range
+ 000.0010 Kohm
- 000.0010 Kohm
16
Short
3 MOhm Range
= + 0.000014 Mohm
= 0.000014 Mohm
17
Short
30 MOhm Range
+ 00.00083 Mohm
- 00.00083 Mohm

Advertisement

Table of Contents
loading

Table of Contents