Dc Current Function - Of; Dc Current Function - HP 3457A Service Manual

Hide thumbs Also See for 3457A:
Table of Contents

Advertisement

Performance Tests
3457A
DC Current Performance Tests Cont'd
3457A
|3457A
| 3457A
90
day
limits
1 year
limits
Input
Range
|Set
Up
High
Low
High
Low
Open
| 300
uA]
DCI
+000.0104
uA
-000.0104
uA
+000.0104
uA
-000.0104
uA
Open
3 mA}
DCI
+0.000104
mA
-0.000104
mA
+0.000104
ma
-0.000104
mA
Open
30 mA}
DCI
+00.00104
mA
-00.00104
ma
+00.00104
mA
-00.00104
mA
Open
|300 mA]
OCI
+000.0204
ma
-000.0204
ma
+000.0204
mA
-000.0204
ma
Open
1A
ocl
+0.000604
A
-0.000604
A
+0.000604
A
-0.000604
A
300
uA}]300
uA}
DCI
300.0704
uA
299.9296
uA
300.1304
uA
299.8696
uA
3 mA
3 mA}
DCI
3.000704
ma
2.999296
mA
3.001304
mA
2.998696
mA
30 mA|
30
mA|
DCI
30.00704
mA
29.99296
mA
30.01304
mA
29.98696
mA
300
mA|300
mA}
DCI
300.2304
mA
299.7696
mA
300.2604
mA
299.7396
mA
1A
1A
ocl
1.001304
A
0.998696
A
1.001304
A
0.998696
A
Table 4-5. DC Current Offset and Gain Test Limits
4-24, DC Current Function - Offset Test
4-25. Equipment Required. This procedure does not require any test equipment.
1. Set the HP 3457A to the DC Current function (DCI), 300 uA range, and the number of digits dis-
played (DIGITS DISP) to six.
2. Use the Test Record to record the current offset readings on the 300 uA, 3 mA, 30 mA, 300 mA
and | A ranges.
3. If any of the offset readings are beyond the limits specified in Table 4-5 and on the Test Record,
the instrument should be calibrated or repaired. Refer to Section V of this manual for calibration
procedures.
4-26. DC Current Function - Gain Test
4-27. Equipment Required.
A DC Current Source capable of providing currents of 300 uA, 3 mA, 30 mA
(+ 007%), 300 mA (+ .026%), and 1 A (+ .04%) is required for the following procedure.
4-8
1. Set the DC Current Source to 300 uA and connect it to the HP 3457A front panel I and LO input
terminals.
2. Set the HP 3457A to the 300 uA range and the number of digits displayed (DIGITS DISP) to six.
3. Use the Test Record to record the full-scale current readings for the 300 uA, 3 mA, 30 mA, 300 mA
and | A ranges.
4. If any
of the full-scale readings are beyond
the limits specified
in Table
4-5 and
on the Test
Record, the instrument should
be calibrated or repaired. Refer to Section V of this manual
for
calibration procedures.
5. Reduce the output of the DC Current Source and disconnect it from the HP 3457A input terminals.