SECTION IV
PERFORMANCE TESTS
4-1. INTRODUCTION
4-2, Section IV contains Performance Tests designed to verify the accuracy of the HP 3457A
Multimeter.
Accuracy
specifications
are
listed
in Table
1-1 of this manual.
This section
also contains
Operational
Verification Tests which
provide a more
rapid method
of testing the operation and accuracy of the unit.
The Operational
Verification Tests are designed to provide a 90% confidence that the HP 3457A
is opera-
tional and meets specifications. Both the Performance Tests and Operational
Verification Tests can be per-
formed without access to the interior of the instrument.
4-3. EQUIPMENT REQUIRED
4-4, Equipment
required
for the performance
tests and
operational
verification
tests are
listed
in the
Recommend
Test Equipment
table in Section
I of this manual.
Equipment
other than that recommended
may be used as long as the critical specifications are met
4-5. TEST RECORD
4-6, Results of the performance or operational verification tests may be tabulated on the Test Record lo-
cated at the end of the respective procedures. The Test Record lists all of the tested specifications and their
acceptable limits. It is suggested that the performance tests or operational
verification tests be performed
and the results tabulated when
the instrument
is received.
These results can be used for comparison
with
periodic calibration results.
4-7. CALIBRATION CYCLE
4-9. TEST CONSIDERATIONS
4-10. General
4-1