HP 3457A Service Manual page 34

Hide thumbs Also See for 3457A:
Table of Contents

Advertisement

Installation
Fuse
for
95
to
132Vac
Operation:
0.2 AT
HP Part Number 2110-0235
Fuse Cap for 95 to 132Vac operation:
Gray
HP
Part
Number
2110-0565
~ LINE
~LINE
48-66 Hz.
30 VA MAX
48-66 Hz.
30 VA MAX
100Vac
Line Selection
120Vac
Line
Selection
Fuse for 209 to 264Vac operation:
.08
AT
HP
Part
Number
2110-0719
Fuse
Cap
for
209
to
264Vac
operation:
Black
HP Part Number 2110-0567
~LINE
~LINE
48-66
Hz.
30 VA MAX
48-66
Hz.
30 VA MAX
220Vac Line Selection
240vVac Line Selection
Figure 2-2, Line Voltage and Fuse Selection
2-5. PREPARATION
FOR USE
2-6. Power Requirements
2-7. The HP 3457A requires a power source of 100, 120, 220, or 240 Vac (+5%, -10%), 48Hz to 66Hz single
phase. Power consumption is less than 30VA.
Before
connecting
power
to the HP 3457A, make
certain
that the line selection
switches (located on the rear panel) have heen set to accept the available power
source and that the proper fuse is installed.
2-8. Line Voltage Selection
2-9, The line voltage selection switches are located on the rear panel below the power
connector and fuse.
Figure 2-2 shows
the appropriate
switch
positions
and
fuse
requirements
for the various
power
line
voltages.
2-10. Power Cords
2-11, This instrument
is equipped
with a three-wire
power
cable.
This cable, when
connected
to an ap-
propriate ac power receptacle, grounds the metal parts of the cabinet. The type of plug supplied with the
power cable depends upon the country of destination.
Figure 2-3 illustrates the power plugs available. The
HP Part Number listed below each figure is the part number for the complete power cord assembly. If the
appropriate power
cord is not supplied
with your
instrument,
notify your nearest
HP Sales and Service
Office and a replacement cable will be supplied.