MANUAL CHANGES MODEL 5355A
(05355-90013)
SERIAL PREFIX OR
SERIAL NUMBER
Page 1-2,
Table
1
All Serials
Page
1~3S Table
1
All Serials
2132A
Page 1-*l9
Table
1
All Serials
2132A
Page 1-5.
General
All Serials
CHANGES
1.
Model 5355A Specifications:
Maximum Input:
>Add two asterisks
(**)
to "+5 dBm" for the 5356C.
>Add the following footnote:
** In Pulse Mode
1.5 to 5.0 GHz,
0 dBm (+5 dBm typical).
>Add two asterisks (**) to the "+15dBm" on all ranges
for the 5356C.
>Change the two-asterisk footnote to read as follows:
»»#* jn pulse Mode,
OdBm (+5dBm typical)."
1.
Model 5355A Specifications:
>Delete the asterisk(*) next to 5355A,
under PULSE MODE
column.
>Delete the Min Pulse Width,
in the 5355A column, to
150ns.
>Delete,
at bottom of page,
"Specifications apply only to
external gating of 53*15/5355".
>Change the MAN MODE specification from "60ns" to "75ns"
in
the area labeled "PULSE MODE"
"Pulse Width" "MIN".
>Change the "PULSE MODE" "Accuracy" specification (lower
right-hand corner)
from " (. 08EXT GATE WIDTH)" to
"(. 12/EXT GATE WIDTH)".
1.
Model 5355A Specifications:
>Change maximum Operating Temperature to 55 degrees.
>Change 40ns to 65ns under DEFINITIONS,
rms jitter=.
Information:
>Add
the following paragraph after Table
1-2:
OPTION W30
(Extended Hardware Support) provides two addi¬
tional years of return-to-HP hardware-service support.
Option W30 is available only at time of purchase.
Service
contracts are available from Hewlett-Packard for instruments
which did not include Option W30 at time of purchase.
For
more information, contact your nearest Hewlett-Packard
Sales and Support office (offices are listed at the back of
this manual).