Performance Tests
Model 3314A
SPECTRUM ANALYZER
TERMINATION
NO CONNECTION
Figure 4-20. Harmonic Distortion Connections
Procedure:
A. Preset the 3314A.
B. Set the 3314A as follows:
ЕТЕППЕН
au caque
cea ЫРА Deka SEX КЫЗ 20Hz
Amplitude
..........................-...
10Vp-p
C. Connect the 50 Ohm Feedthrough Termination to
the 3314A's
Main
Output.
Connect
the 5000
Feed-
through Termination to the 3314A's X3 Output.
D. Connect the spectrum analyzer's IMQ input to the
3314A's X3 Output. Set the spectrum analyzer to measure
the fundamental and at least seven harmonics. Use the
Marker function to verify that the level of the largest har-
monic is more than 53dB below the fundamental.
E. Repeat step D for 3314A frequency settings of
50kHz and 1MHz. The largest harmonic should be more
than 53dB below the fundamental at 50kHz, and more
than 38dB below at ІМН2.
од
5002
FEEDTHRU
4-22. SQUARE WAVE RISE/FALL TIME
This test uses a wide bandwidth oscilloscope to measure
the rise and fall time of a IOKHz, 30Vp-p square wave.
Required Equipment:
Oscilloscope
f
50 Ohm Feedthrough Termination
A 5000 Feedthrough Termination consisting of:
1 4990 resistor, -hp- part number 0698-4123
Procedure:
A. Preset the 3314A.
B. Set the 3314A as follows:
оо
noces
зеш
uw Us а
RES
ЫЗ
IOkHz
Атар
nde scd ditus
Tip
ba eee а
10Vp-p
РОШ
voee
xe
rv
ees EVE
Square Wave
FEEDTHRU
TERMINATION
NO CONNECTION
Figure 4-21. Square Wave Rise/Fall Time Connections