Hitachi TM3000 Instruction Manual page 260

Tabletop microscope
Table of Contents

Advertisement

Code
Error Message
1575
The power supply of +18V for
C2 was shut down.
Start up the main unit and
application again.
1576
The power supply of +18V for
C1 was shut down.
Start up the main unit and
application again.
1577
The power supply of +18V for
OBJ was shut down.
Start up the main unit and
application again.
1578
The power supply of -15V for
STG was shut down.
Start up the main unit and
application again.
1579
The power supply of +15V for
STG was shut down.
Start up the main unit and
application again.
1580
The power supply of -15V for
DEF was shut down.
Start up the main unit and
application again.
Countermeasure
An overcurrent in the power supply or a blown
fuse (F14 on the COLHVC00 board) was found.
Click the [OK] button to dismiss the message.
Restart the main unit. If the message appears
repeatedly, contact the nearest service
representative.
An overcurrent in the power supply or a blown
fuse (F13 on the COLHVC00 board) was found.
Click the [OK] button to dismiss the message.
Restart the main unit. If the message appears
repeatedly, contact the nearest service
representative.
An overcurrent in the power supply or a blown
fuse (F12 on the COLHVC00 board) was found.
Click the [OK] button to dismiss the message.
Restart the main unit. If the message appears
repeatedly, contact the nearest service
representative.
An overcurrent in the power supply or a blown
fuse (F11 on the COLHVC00 board) was found.
Click the [OK] button to dismiss the message.
Restart the main unit. If the message appears
repeatedly, contact the nearest service
representative.
An overcurrent in the power supply or a blown
fuse (F10 on the COLHVC00 board) was found.
Click the [OK] button to dismiss the message.
Restart the main unit. If the message appears
repeatedly, contact the nearest service
representative.
An overcurrent in the power supply or a blown
fuse (F9 on the COLHVC00 board) was found.
Click the [OK] button to dismiss the message.
Restart the main unit. If the message appears
repeatedly, contact the nearest service
representative.
5 - 43

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents