5-9.

Adjustment Procedures

5-10.
The following paragraphs contain the adjustment procedures for the HP5501B Laser Head. A
performance test is given in

5-11. Reference Temperature Adjustments

5-12.
This adjustment sets the laser preheat temperature to 105°C (221°F). The adjustment must be
performed when the laser tube is at room temperature, 21 to 25°C (69.8 to 77.0°F).
5-13.
Test Equipment Required. The reference temperature adjustment requires the following test
equipment: Digital Voltmeter, HP Model 3435A

5-14. Preliminary Procedure

5-15.
This adjustment must be made with the laser tube at room temperature. If the laser High Voltage
Power Supply (A2) or heater has been on, allow at least two hours with ±15 Volts off before
performing this procedure. Regardless of whether power is on or off, observe the safety
precautions in paragraph 5-4 and proceed as follows:
Remove the HP5501B Laser Head covers by following the procedure outlined in paragraph 8-55.
a.
Check that the TEST-NORM switch (A1S2) located on the A1 Connector Board is set to
NORM. See Figure 5-1.

Figure 5-1. A1 Connector Board, TEST-NORM Switch and Adjustment Locations

5-2
Section
4.
HP5501B
Adjustments