Figure 8-11. Factory-Adjusted Sampler Mount (Mp1) - HP 5501B Operating And Service Manual

Laser head
Table of Contents

Advertisement

Sampler Mount (MP1) is adjusted at the factory to ensure that the laser
beam is not truncated as it exits the laser head. Two screws (H5) secure
the Sampler Mount to the shield (MP10).
8-60.
To remove the A5 Laser Assembly (HP Part No. 05501-60102) from the HP5501B chassis,
perform the following procedure.
a.
Remove ±15 Volt power from HP5501B Laser Head.
b.
Remove front panel and side covers as described in paragraph 8-55.
c.
Disconnect laser assembly connector A5P1 from A1J4.
d.
Remove two Pozidriv screws (H7) that secures rear panel (MP4) to laser base (MP2).
8-18

Figure 8-11. Factory-adjusted Sampler Mount (MP1)