16. Wait for the readings to stabilize > Press
NOTE: If the trial weight was NOT of sufficient mass to affect the
imbalance, the instrument will display a warning prompt. You can
redo the trial reading several times if required, or continue with
balancing.
17. Press the appropriate key when the instrument displays the prompt
Will the Trial Weight be removed now?
NOTE: You can redo the trial reading several times if required or
continue with balancing.
Section 12 — Balancing Walkthrough