ABB 4690 User Manual page 30

4690 series. turbidity systems. accurate, reliable turbidity measurement
Hide thumbs Also See for 4690:
Table of Contents

Advertisement

4690 Series
Turbidity systems
Continued from previous page
1 3
Clean Int.
(Hr)
– – – – –
Test Cleaner
NO
YES
– – – – –
Bubble Rej.
NO
YES
2 0
Bubble Period
Return to
top of page
2 0
Signal Filter
- - - - -
SET UP OUTPUTS
28
Cleaning Interval
Note. Displayed only if
Sensor Type
Using the
and
keys, select the cleaning interval required. Values available are: 15 mins,
30 mins, 45 mins, 1 hr, then 1 hour increments to a maximum of 24 hours.
Test Cleaner
Note. Displayed only if
Sensor Type
Select YES to activate the wiper unit diagnostics function and enable visual indication of wiper unit
function.
Select No to deactivate the wiper unit diagnostics function.
Bubble Rejection
Use in applications where degassing of the sample takes place.
When set to YES, the analyzer ignores short duration spikes in the turbidity reading caused by the
formation of small bubbles in the flowcell due to degassing.
Bubble Period
Note. Displayed only if
Bubble Rej.
Using the
and
keys, adjust the value to between 2 and 60 seconds until short duration
spikes are no longer seen on the display.
Note. A value of 20 seconds is normally sufficient for most applications.
Signal Filter
Use in applications where the turbidity readings are noisy.
Using the
and
keys, adjust the value to between 4 and 180 seconds.
is set to YES, set a value of 4 seconds for optimum performance.
Note. If
Bubble Rej.
Advance to
page – see Section 8.4, page 29.
Set Up Outputs
8 Configuration and Electrical Calibration
is set to 7998-001, 7998-002, 7998-011 or 7998-012
is set to 7998-001, 7998-002, 7998-011 or 7998-012
is set to YES
IM/4690–EN Rev. A