INCUBATOR COMPONENTS
1. Outer door: Sticks to frame with magnetic seal. Door heater is installed in the door panel. The
door opening is reversible. Contact Sanyo representative or agent to change the door hinge from left to
right or vice versa.
2. Inner door: Made of tempered glass, however avoid excessive impact on the glass.
3. Leveling foot: Screw type for adjusting the height. Adjust the foot so that the unit can be level.
4. Tray: Can be pulled toward you.
5. Tray support: Can be removed by lifting the front side and pulling toward you.
6. Side support: Right and left side supports can be removed for disinfection. See page 27 and 28.
7. Top duct: Inlet of circulating air. Removable.
8. Rear duct: Flow path for circulating air. Removable.
9. Fan (inside the rear duct): Made from polypropylene resin. Can be disinfected by an autoclave.
10. Sample air outlet: This also functions as an internal gas outlet, so do not cap it.
11. Connecting port for CO
gas pipe (rear side): When an optional component MCO-21GC (gas
2
cylinder changeover accessory) is installed, both A and B are available. If MCO-21GC is not used, only
A is available. Refer to page 17 for gas cylinder connection. Ensure that the gas pressure is set at
2
0.03MPaG (0.3kgf/cm
G, 4.3psiG). Refer to page 18 for automatic cylinder changeover.
12. Door switch: Detects the door opening/closing and stops the circulating fan and electromagnetic
valve for CO
when door is open. UV lamp is also deactivated by door opening (When an optional UV
2
system kit MCO-18UVS is installed).
13. Humidifying pan: Use sterile distilled water to fill the pan. Install the pan properly so that it can be
covered with the pan cover.
14. Humidifying pan cover: Prevents UV light being exposed to the chamber. When filling the pan,
lift the front side and take out the pan. See page 29 for details.
15. UV lamp (When an optional UV system kit MCO-18UVS is installed):
Sanyo UV lamp does not generate ozone. Never look at the UV light directly. For replacement, contact
Sanyo representative or agent.
16. Water level sensor for humidifying pan: Detects the water level in the humidifying pan. See
page 23 for details.
17. Remote alarm terminal: Refer to page 13.
18. Access port: When not in use, cap with the rubber cap on both outside and inside.
19. Power switch: Main switch of the unit. Also
functions as an over-current breaker. The switch is
covered by a switch cover to prevent the accidental
push. To turn on or off the switch, remove the switch
cover by loosening the screw.
See figure on the
right.
Switch cover
20. Glow starter (When an optional UV system kit
MCO-18UVS is installed):
For UV lamp (model; FG-7P)
57