ROUTINE MAINTENANCE
9. To removing the circulating fan, pull out the spring and then pull the fan. See Fig. 7.
10. Loosen and remove the screw fixing the clamp and remove the clamp. See Fig. 8 and Fig. 9.
11. By lifting the side support, take it off from the hooks. See Fig. 10.
Fig. 7
Fig. 8
Fig. 9
Fig. 10
12. Replace all attachments in the chamber with the reversed order mentioned above.
Note:
• Always insert the fan on the motor shaft surely.
Check that the fan does not hit the rear panel by
turning the fan manually. Improper insertion may
cause poor performance.
• As shown in the figure, set the shelf with the edge
bent downwardly positioned at the front. Improper
setting cause tilted or unstable condition.
Downward bend