ROUTINE MAINTENANCE

Removal of attachments

Note:
Take care not to damage the UV lamp (when an optional UV system kit MCO-18UVS is installed) or
water level sensor at the time of removal or replacement of attachments.
1. Close the valve of the gas cylinder and turn off the power.
2. Open the outer door. Remove 2 pins fixing the inner hinge and remove the inner door.
3. Pull out the all trays. See Fig. 1
4. Remove all tray supports on the right and left side by lifting up the front side. See Fig. 2.
5. By lifting up the humidifying pan cover, unhook it at the rear and remove the cover as shown in Fig. 3.
6. Take out the humidifying pan. See Fig. 4
7. After removing 2 screws fixing the top duct, take out the top duct by removing pin. See Fig. 5.
8. To remove the rear duct, lift it up as shown in Fig. 6.
Fig. 1
Fig. 4
73
Fig. 2
Fig. 5
Fig. 3
Fig. 6