HP 35601A Operating And Service Manual page 73

Spectrum analyzer interface
Table of Contents

Advertisement

6 1 9 . Output Path to 3582/3585
This test checks the circuits in the signal path shown in Figure 6-11. If the previous tests have
been performed, the only circuit being checked that hasn't been tested, is switch S5 A
voltage is generated by the Digital to Analog Converter. The voltage is then measured by an
-hp- 3455A Voltmeter at the IMO OUTPUT to 3585A and at the Channel B Output In this
test, the Adaptive Coupler is in the DC mode (L2=0). At one point in the program, the user
is given the option of discontinuing the test if it appears to be failing. If he chooses to con­
tinue the program, the test is completed and the results are printed.
Initializing the Test
With the High Frequency Menu displayed, press shift SFK Kl (-hp-9836) or SFK Kl 1 (-hn-
9845B) to initialize the test. After initialization, follow the instructions displayed on the
computer,
What if the Test Fails?
sf^SSLJSS^
leStS h a v e p a s s e d
' **
o n l v
P
r o b a b l e rause
of failure is the switch
f
P i SWITCH" subprogram can be used to open and close this switch when
troubleshooting. Refer to Service Group Three for schematics and component locators
Equipment Set Up
O
35601
O O
O
O °
1 MQ Signal Path Measurement
O
35601
O O
o o °
{OUTPUT
I TO 3582 B
-hp- 3582A Channel B Signal Path Measurement
Finare 6-11, Channel B and tflSO Outputs
Channel B and 1MQ Outputs,
6-27

Advertisement

Table of Contents
loading

Table of Contents