8. Insert a "T" from the tool kit in between the regulator and the clear oxygen tubing and connect the
pressure gauge.
9. Turn the device on and set the flow meter to 5 LPM (1 LPM for units equipped with pediatric
flow meters).
10. With the gauge held in a vertical position, verify that the regulated pressure is at 5.5 ± 0.25 psig.
11. If the regulated pressure needs adjusted, place a 3/32" Allen wrench in the adjusting screw on the
end of the regulator.
If the regulated pressure is low, turn the wrench clockwise to increase the regulated pressure. If the
regulated pressure is high, turn the wrench counter-clockwise.
12. Disconnect the pressure gauge and "T" fitting.
13. Reconnect the clear oxygen tubing to the pressure regulator.
14. Install the front and rear cabinets.